Patient Database:
The examiner can enter new patient information either immediately before the test, or as a batch, before anyone is tested.
Parameter Window:
The following parameters may be modified:
frequency range
stimulus intensity
F2/F1 ratio
DP definition
DPs per octave
averaging time per frequency
pass/refer criteria
Artifact Rejection:
User definable criteria and stopping rules.
Adjustable Parameters: Range: 500-16,000 Hz
Number of test points per octave - 0-50
Intensity: f1 level (0-70 dB) 12 level (0-70 dB)
Minimum number of valid test points to pass a test
Minimum DP level to pass
Ratio: 1.1 to 1.8 F2/F1
Ratio: 1.1 to 1.8 F2/F1
Averaging time: 0.25 to 20 sec DP definition points: f2-f1; 2f1-f1; 2f1-f2; 3f1-2f1; 3f2-2f1; 4f1-3f2,; 4f2-3f1
Artifact Rejection Criteria:
Noise level for rejection
Noise level above calibrated
Noise floor rejection
Stopping Rules Criteria:
Noise Floor level for accepting DP measurement
Acceptable DP to Noise Floor ratio
Maximum Signal Duration for a frequency point
Computer System Recommendations:
IBM TM compatible portable computer having an available PCMCIA type II slot
8 MB of RAM
75 MHz or faster in an 486 DX or Pentium TM machine
High Density (1.44) floppy disk drive
Color Monitor Windows 98TM or 95TM Operating System
1 GB Hard Drive
System Requirements:
Environmental conditions: operating temperature must be within the range of +59°F to +95°F or +15C to +35°C Relative humidity must be within the range of 30 to 90%
Weight: Less than 6 oz (without computer)
Software Ear tips: 8 different disposable ear tip sizes
Upgradeable |